Beam Qualities

Home / Calibration / Calibration Types / Beam Qualities

AAPM Accredited NIST-Traceable Beam Qualities

Therapy Ion Chamber

Diagnostic X-Ray Ion Chamber

60Co (in water per TG-51 or air kerma per TG-21)

Choose from the Tungsten (radiographic) anode beams or the Molybdenum (mammographic) anode beams listed below

137Cs

X-Ray (choose from the Tungsten anode beams listed below)

Tungsten Anode X-Ray Beam Qualities

Beam Code

Added Filter (mm Al or Cu)

HVL 1 (mm Al)

HVL 1 (mm Cu)

HC (Al)

Air Kerma Rate (mGy/sec)

UW30-L

0.24 Al

0.24

61

2.295

UW40-L

0.46 Al

0.51

59

2.114

UW50-L

0.63 Al

0.80

61

2.882

UW80-L

1.20 Al

1.82

60

3.601

UW100-L

1.80 Al

2.81

59

3.930

UW20-M

0.19 Al

0.15

79

0.869

UW30-M

0.41 Al

0.36

66

1.308

UW40-M

0.77 Al

0.75

68

1.291

UW50-M

0.92 Al

1.03

66

2.105

UW60-M

1.66 Al

1.68

68

1.758

UW80-M

3.09 Al

3.07

68

1.737

UW100-M

5.21 Al

5.10

73

1.746

UW120-M

2.21 Al + 0.13 Cu

6.78

76

2.417

UW150-M

2.49 Al + 0.31 Cu

10.30

0.66

86

2.037

UW200-M

0.89 Al + 1.06 Cu

14.73

1.62

94

1.892

UW250-M

4.29 Al + 2.86 Cu

18.49

3.13

98

1.431

Molybdenum AnodeX-Ray Beam Qualities

UW Beam Code

1st HVL (mm Al)

Added Filter (mm)

Air Kerma Rate (mGy/sec)

Mo

Rh

UW23-MO

0.280

0.032

0.568

UW25-MO

0.305

0.032

0.733

UW28-MO

0.336

0.032

0.992

UW30-MO

0.359

0.032

 1.243

UW35-MO

0.390

0.032

1.795

UW28-MR

0.424

0.03

0.673

UW32-MR

0.453

0.03

 1.023

UWMRRC